FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3151710 · Received June 7, 2013

Report

Report Number
2182208-2013-01281
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
March 22, 2013
Report Date
March 22, 2013
Manufacturer
RICE CREEK MFG
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: PRODUCT ID 2067L RADIO FREQUENCY PROGRAMMER HEAD. (B)(6). (B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS CONFIRMED INTERMITTENT INTERROGATION, EVEN WITH A KNOWN GOOD RADIO (RF) FREQUENCY PROGRAMMER HEAD. LOOSE RF AND ELECTROCARDIOGRAM (ECG) CONNECTORS WERE FOUND ON THE LINK ELECTRONIC MODULE (LEM) BOARD, WHICH WAS REPLACED AND CALIBRATED. ANALYSIS ALSO FOUND THAT THE DISPLAY LOWERED BY ITSELF IN SOME POSITIONS AND THE LOWER LEFT AND LOWER RIGHT HINGES WERE REPLACED. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE ATTEMPTING INTERROGATION, THE PROGRAMMER WAS UNABLE TO RECOGNIZE IMPLANTABLE HEART DEVICES. THE PROGRAMMER HEAD WAS CHANGED OUT BUT THIS DID NOT RESOLVE THE ISSUE. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE ATTEMPTING INTERROGATION, THE PROGRAMMER WAS UNABLE TO RECOGNIZE IMPLANTABLE HEART DEVICES. THE PROGRAMMER HEAD WAS CHANGED OUT BUT THIS DID NOT RESOLVE THE ISSUE. THE PROGRAMMER WAS RETURNED FOR SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252371 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY RICE CREEK MFG 2090

Patients

Seq Age Sex Outcome Treatment
1