FDA Adverse Event Malfunction Summary report: N

CARELINK

MDR report key: 3151674 · Received June 7, 2013

Report

Report Number
2182208-2013-01284
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
MEDTRONIC, INC.
Product Code
DXY
PMA / PMN Number
P890003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: ANALYSIS FOUND THAT THE PROGRAMMER WAS ABLE TO INTERROGATE USING A KNOWN GOOD HEAD AND MOVING THE CONNECTION DID NOT MAKE ANY CHANGES, PROGRAMMER PASSED FUNCTIONAL TESTING.

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. CONCOMITANT PRODUCT: PRODUCT ID 2067L RF HEAD. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED DISPITE CHANGING PROGRAMMING HEADS, THE PROGRAMMER HAS ON / OFF ISSUES DETECTING DEVICES. THE PROGRAMMER WAS RETURNED TO CHECK THE CONNECTION PORT ON THE PROGRAMMER. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252508 CARELINK PULSE-GENERATOR, PACEMAKER, IMPLANTABLE DXY MEDTRONIC, INC. 2090W

Patients

Seq Age Sex Outcome Treatment
1