FDA Adverse Event Injury Summary report: N

VIRTUOSO DR

MDR report key: 3151620 · Received June 7, 2013

Report

Report Number
3004209178-2013-08832
Event Type
Injury
Date Received
June 7, 2013
Report Date
March 1, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
LWS
PMA / PMN Number
P980016
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS REVEALED THE RETURNED DEVICE FOUND A HIGH CURRENT DRAIN CONDITION DUE TO CURRENT LEAKAGE IN A CERAMIC CAPACITOR. THE DEVICE MET 63% OF THE EXPECTED LONGEVITY. A 5076, IMPLANTABLE PACING LEAD, IMPLANTED: (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR  (ICD) EXPERIENCED PREMATURE BATTERY DEPLETION. THE SYSTEM WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
252087 VIRTUOSO DR DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MEDTRONIC MED REL MEDTRONIC PUERTO RICO D154AWG

Patients

Seq Age Sex Outcome Treatment
1 00056 YR Hospitalization| R 6947, IMPLANTABLE DEFIBRILLATION LEAD