PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
Report
- Report Number
- 2183613-2013-00506
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- February 27, 2013
- Report Date
- March 5, 2013
- Manufacturer
- MEDTRONIC MILACA, INC.
- Product Code
- LWP
- PMA / PMN Number
- P820003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
PRODUCT EVENT SUMMARY: ANALYSIS COULD NOT CONFIRM THE REPORTED EVENT. HOWEVER, IT WAS NOTED THAT THE UPPER CASE WAS BROKEN AND CONTAMINATED, THE LOWER CASE WAS BROKEN AND CONTAMINATED, THE BATTERY RELEASE WAS CONTAMINATED, TWO SIDE BAIL COVERS WERE BROKEN, THE LEAD FLEX COVER WAS CONTAMINATED, THE BATTERY CONTACTS WERE COMPRESSED, THE RING WAS BENT, THE BATTERY DRAWER WAS BROKEN, AND THE DISPLAY WAS OUT OF SPECIFICATION WITH PIXEL SEGMENTS MISSING.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THE DEVICE IS LOCKED UP. THE DEVICE WAS RETURNED FOR REPAIR. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 252327 | PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE | LWP | MEDTRONIC MILACA, INC. | 5388 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |