FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3151413 · Received May 30, 2013

Report

Report Number
1720753-2013-06539
Event Type
Malfunction
Date Received
May 30, 2013
Date of Event
May 9, 2013
Report Date
May 30, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SYSTEMS INTERFACE BOARD WAS REPLACED AND THE SOFTWARE WAS RELOADED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED INTERMITTENT COMMUNICATION FAILURE ERROR MESSAGES. THIS ERROR MESSAGE LIKELY CAUSED THE SYSTEM TO LOCK-UP OR PREVENTED THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237817 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1