FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 3151375 · Received May 30, 2013

Report

Report Number
1720753-2013-06572
Event Type
Malfunction
Date Received
May 30, 2013
Date of Event
May 13, 2013
Report Date
May 30, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE SENSOR WAS ADJUSTED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM DISPLAYED A COMMUNICATION ERROR. A COMMUNICATION FAIL ERROR MESSAGE WILL LIKELY RESULT IN A SYSTEM LOCKUP, NO BOOT OR SHUT DOWN SITUATION. THERE ARE NO REPORTS OF PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
237776 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1