HARMONIC ACE CURVED SHEARS INSTRUMENT
Report
- Report Number
- 2955842-2013-02040
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- April 17, 2013
- Report Date
- May 15, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K112584
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT WITH THE FOLLOWING CLARIFICATION: THE INSERT FAILED A SELF-TEST ON THE GENERATOR. A FUNCTIONAL ENERGY TEST WAS PERFORMED USING A HARMONIC GENERATOR AND HANDPIECE. THE INSERT FAILED A SELF-TEST, CAUSING THE GENERATOR TO PRODUCE AN INSTRUMENT ERROR MESSAGE. A HIGH PITCHED NOISE WAS HEARD DURING THE SELF TEST. THE INSERT WAS VISUALLY INSPECTED UNDER MAGNIFICATION AND A HAIRLINE CRACK WAS FOUND ON THE BLADE. THE CRACK WAS LOCATED .400 BELOW THE DISTAL TIP. A GOUGE AND SCRATCH WAS ALSO OBSERVED ON THE BLADE EDGE IN THE SAME AREA AS THE CRACK. ENGINEERING CONCLUDED BASED ON LOCATION OF DAMAGE, THE GOUGE WAS LIKELY CAUSED BY MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI MYOMECTOMY PROCEDURE THE HARMONIC ACE CURVED SHEARS INSERT WORKED FOR ABOUT THE 20 MINUTES AND THEN THE GENERATOR WOULD NOT ENGAGED. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253311 | HARMONIC ACE CURVED SHEARS INSTRUMENT | ENDOSCOPIC INSTRUMENT ACCESSORY | NAY | INTUITIVE SURGICAL,INC. | 400272-02 | M10121022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |