COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER
Report
- Report Number
- 1061932-2013-01016
- Event Type
- Malfunction
- Date Received
- June 6, 2013
- Date of Event
- May 10, 2013
- Report Date
- May 12, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K010765
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED AND NOTICED EVIDENCE OF A DRIED LEAK ON THE CBC (COMPLETE BLOOD COUNT) SOLENOIDS, PELTIER ASSEMBLY, AND PUMP MODULE. THE FSE DISCOVERED A DISCONNECTED TUBING FROM VACUUM CHAMBER VC2 TO SWEEPFLOW ASSEMBLY. THE FSE REPLACED ALL PILOT ACTUATORS ON THE PUMP MODULE, THE RESTRICTOR TUBING AND THE CHECK VALVE TO VC2. THE FSE NOTED THAT THE COMPONENTS THAT HAD DRIED LEAK WOULD BE REPLACED AT A LATER DATE AS A PRECAUTION. THE FSE ALSO OBSERVED THAT THE FRONT AND REAR BLOOD DETECTORS WERE ABOVE THE ACCEPTABLE UPPER VOLTAGE LIMIT. THE FSE REMOVED AND CLEANED THE BSV (BLOOD SAMPLING VALVE) ASSEMBLY AND ALIGNMENT BAR, AND ADJUSTED THE BLOOD DETECTORS TO ESTABLISHED VOLTAGE VALUES. THE FSE ALSO REPORTED THAT THE % CV (COEFFICIENT OF VARIATION) VALUES FOR DIFFERENTIAL AND RETIC SCATTER WERE ON THE BORDERLINE OF THE ESTABLISHED VALUE. THE FSE RESOLVED THE ISSUE BY CLEANING THE FLOWCELL AND ADJUSTING THE TTM (TRIPLE TRANSDUCER MODULE) ALIGNMENT. FAILURE MODE OF THE EVENT IS ATTRIBUTED TO A DISCONNECTED TUBING FROM VC2 TO SWEEPFLOW ASSEMBLY, TO THE FRONT AND REAR BLOOD DETECTORS, AND TO THE FLOW CELL MODULE. THERE WAS NO INJURY OR EXPOSURE REPORTED BY THE CUSTOMER AS A RESULT OF THE LEAK. (B)(4).
THE CUSTOMER REPORTED THAT THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER FAILED STARTUP FOR RBC (RED BLOOD CELL) AND PLATELETS. NO ERRONEOUS RESULTS WERE ASSOCIATED WITH THE EVENT AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251538 | COULTER® HMX HEMATOLOGY ANALYZER WITH AUTOLOADER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |