RESERVOIR 3ML
Report
- Report Number
- 3004209178-2013-94456
- Event Type
- Malfunction
- Date Received
- June 6, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
INSPECTED TWO SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER INSPECTION. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER INSPECTION. NO LEAKS OR O-RINGS DEFECTS WERE FOUND DURING TEST. RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY.INSPECTED ONE OPENED AND USED RESERVOIR. PERFORM MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER INSPECTION. NO LEAKAGE ANOMALY WAS OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. RESERVOIR CONNECTED AND LOCKED IN PLACE PROPERLY.
IT WAS REPORTED THAT INSULIN FROM THE RESERVOIR LEAKED INTO THE RESERVOIR COMPARTMENT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 251354 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A | H8476270 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |