FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3150909 · Received June 6, 2013

Report

Report Number
3004209178-2013-94491
Event Type
Injury
Date Received
June 6, 2013
Date of Event
May 21, 2013
Report Date
May 21, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. UNABLE TO PERFORM BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME/FILL ANOMALY. HOWEVER, THE INSULIN PUMP PASSED THE DISPLACEMENT TEST AND BOLUS DELIVERY. NO MOTOR ERROR ALARMS OCCURRED DURING TEST. MOTOR TEST PASSED. THE INSULIN PUMP WAS RETURNED WITH A BROKEN RESERVOIR TUBE LIP, MISSING O-RING RESERVOIR TUBE, CRACKED RESERVOIR TUBE AND A MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF 525MG/DL. THE CALLER STATED THAT THE DEVICE ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP APPEARS TO BE NORMAL. ASSISTED THE CUSTOMER TO RUN A SELF TEST AND PASSED. THE DISPLACEMENT TEST COULD NOT BE PERFORMED AS CUSTOMER DID NOT HAVE ANY SUPPLIES. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249750 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522NAL

Patients

Seq Age Sex Outcome Treatment
1 39 YR Hospitalization