FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3150895 · Received June 6, 2013

Report

Report Number
3004209178-2013-94465
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 20, 2013
Report Date
May 20, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

INSPECTED THREE RANDOMLY SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER INSPECTION. NO LEAK OR O-RINGS DEFECTS WERE FOUND DURING TEST.INSPECTED FOUR OPENED AND USED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER INSPECTION. NO LEAKAGE ANOMALIES WERE OBSERVED DURING ANALYSIS. CHECKED O-RINGS FOR DEFECTS AND NONE WERE FOUND. ALL RESERVOIRS CONNECTED AND LOCKED IN PLACE PROPERLY.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN FROM THE BOTTOM OF THE RESERVOIR LEAKED INTO THE RESERVOIR COMPARTMENT. THE BLOOD GLUCOSE READING WAS 107MG/DL. NO FURTHER INFORMATION WAS PROVIDED.