FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3150862 · Received June 6, 2013

Report

Report Number
3004209178-2013-94475
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 20, 2013
Report Date
May 20, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH PROTRUDED/LOOSE DRIVE SUPPORT DISK, SEVERELY CRACKED CASE AT THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE, AND MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 225MG/DL. THE CALLER STATED THAT INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS, AND THE INSULIN PUMP ALARMED. THE CUSTOMER MENTIONED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED AND STICKING OUT. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251137 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 43 YR