FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3150821 · Received June 6, 2013

Report

Report Number
2955842-2013-02024
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
December 12, 2012
Report Date
May 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
420309-01
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS UNDAMAGED. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. THERE WERE INDENTATIONS AT THE EDGE OF THE DISTAL PULLEY. THERE WERE VISIBLE SCRATCHES ON THE SURFACE. THE EVIDENCE IS INCONCLUSIVE BUT THE DAMAGE WAS MOST LIKELY DUE TO MISHANDLING. ENGINEERING ALSO FOUND THE DISTAL END OF THE MAIN TUBE HAD A SCRATCH MARK SHOWING LIGHT MATERIAL REMOVAL. THE SCRATCH WAS SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE MAY HAVE BEEN DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S SACROCOLPOPEXY PROCEDURE THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT WIRE WAS NOTED TO BE EXPOSED AND BROKEN. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
250886 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120508 652

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI S SYSTEM, INSTRUMENTS & ACCESSORIES