CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS
Report
- Report Number
- 2015691-2013-20273
- Event Type
- Injury
- Date Received
- June 6, 2013
- Date of Event
- April 23, 2013
- Report Date
- May 10, 2013
- Manufacturer
- EDWARDS LIFESCIENCES
- Product Code
- DYE
- PMA / PMN Number
- P860057/S018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
UNFORTUNATELY, THE EXPLANTED DEVICE WAS NOT RETURNED TO EDWARDS FOR ANALYSIS; THEREFORE, THE SOURCE OF THE REPORTED CALCIFICATION COULD NOT BE ASSESSED. ALTHOUGH THE ROOT CAUSE OF THIS EVENT CANNOT BE CONFIRMED, IT APPEARS THAT THE CALCIFICATION LIKELY CAUSED THE PATIENT'S STENOSIS. CALCIFICATION IS A WELL RECOGNIZED FAILURE MODE OF BIOPROSTHETIC VALVES. THE MECHANISMS FOR BIOPROSTHETIC HEART VALVE TISSUE CALCIFICATION ARE NOT FULLY UNDERSTOOD. MANY FACTORS CAN CONTRIBUTE TO THE ONSET AND PROPAGATION OF CALCIFICATION INCLUDING PATIENT RELATED (E.G. PATIENT AGE, DISEASE STATE, IMMUNE STATUS, AND OTHER CO-MORBIDITIES), PHARMACOLOGICAL, AND INTRINSIC PROPERTIES OF THE VALVE ITSELF. IT IS WIDELY UNDERSTOOD THAT PATIENTS WITH CHRONIC RENAL DISEASE AND PRIOR HISTORY OF CALCIFIC STENOSIS OF THE NATIVE VALVE MAY BE PREDISPOSED TO BIOPROSTHETIC CALCIFICATION. NO FURTHER ACTIONS ARE POSSIBLE WITH THE AVAILABLE INFORMATION.
EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE. IN THIS CASE, IT WAS REPORTED THAT THE VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF APPROXIMATELY 8 YEARS DUE TO SEVERE SYMPTOMATIC PROSTHETIC AORTIC VALVE STENOSIS. PER THE OP REPORT, THE PATIENT HAD NO SIGNIFICANT CORONARY DISEASE. IT WAS NOTED THAT ON THE PREOP CARDIAC CATHETERIZATION THERE WAS AN ORIGIN AT THE LEFT MAIN AT THE LEVEL OF THE AORTIC ANNULUS. DURING EXPLANTATION, THE VALVE WAS HEAVILY STENOTIC, CALCIFIED AND HAD RESTRICTED LEAFLET MOTION. AFTER DEBRIDING THE ANNULUS, IT BECAME CLEAR THAT SEATING OF A NEW VALVE WOULD BE DIFFICULT DUE TO THE PROXIMITY TO THE LEFT MAIN. THE ANNULUS WAS RECONSTRUCTED AND A 19 MM VALVE WAS SEATED WITHOUT DIFFICULTY. THE PATIENT WAS WEANED FROM CARDIOPULMONARY BYPASS WITHOUT DIFFICULTY. THE POST-PUMP ECHO SHOWED SEVERE LVH, BUT NORMALLY FUNCTION LEFT VENTRICLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 250348 | CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS | REPLACEMENT HEART VALVE | DYE | EDWARDS LIFESCIENCES | 3000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Hospitalization| R |