FDA Adverse Event Injury Summary report: N

TALENT AAA

MDR report key: 3150724 · Received June 6, 2013

Report

Report Number
2953200-2013-01084
Event Type
Injury
Date Received
June 6, 2013
Date of Event
May 1, 2013
Report Date
May 16, 2013
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P070027
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: STENT GRAFT MIGRATION, ENDOLEAK, ANEURYSM RUPTURE. DISEASE PROGRESSION. CONCLUSION: STENT GRAFT MIGRATION, ENDOLEAK, ANEURYSM RUPTURE. DISEASE PROGRESSION.

Description of Event or Problem · 1

A TALENT STENT GRAFT WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. ANEURYSM AND VESSEL MORPHOLOGY FROM THE TIME OF IMPLANT ARE UNKNOWN. THE PATIENT DID NOT RETURN FOR ANY FOLLOW-UP VISITS SINCE THE TIME OF IMPLANT. IT WAS REPORTED THAT ON AN UNKNOWN DATE THE PATIENT PRESENTED TO THE HOSPITAL WITH A CONTAINED RUPTURED ANEURYSM. THE AORTIC NECK WAS FOUND TO HAVE DILATED TO 40 MM IN DIAMETER. THIS CAUSED THE BIFURCATED STENT GRAFT TO MIGRATE INTO THE ANEURYSM SAC WITH A PROXIMAL TYPE I ENDOLEAK PRESENT. THE DECISION WAS MADE TO EXPLANT THE TOP PORTION OF THE STENT GRAFT AND IT WAS REPLACED WITH A DACRON GRAFT WHICH WAS SEWN ONTO THE BODY OF THE STENT GRAFT AND TO THE AORTA. THE REMAINDER OF THE BIFURCATED STENT GRAFT AND THE ILIAC LIMBS WERE INTACT AND WERE LEFT IN PLACE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249870 TALENT AAA SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA V00274891

Patients

Seq Age Sex Outcome Treatment
1 00076 YR Required Intervention