FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3150717 · Received June 6, 2013

Report

Report Number
2024168-2013-03541
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 13, 2013
Report Date
May 14, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. SHAFT SEPARATION WAS CONFIRMED. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER UNPACKAGING OF THE 2.5X33 RX XIENCE XPEDITION STENT DELIVERY SYSTEM, INSPECTION WAS PERFORMED PRIOR TO PREPARATION OF THE DEVICE FOR USE. THE CATHETER SHAFT WAS NOTED TO BE SEPARATED INTO TWO PIECES. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. ANOTHER SAME DEVICE WAS PREPPED AND USED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251234 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 2101561

Patients

Seq Age Sex Outcome Treatment
1