FDA Adverse Event
Injury
Summary report: N
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
MDR report key: 3150630
·
Received June 6, 2013
Report
- Report Number
- 1416980-2013-14509
- Event Type
- Injury
- Date Received
- June 6, 2013
- Date of Event
- May 14, 2013
- Report Date
- May 14, 2013
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- KDJ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CAUSE OF THIS PERITONITIS WAS USE ERROR. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.
Description of Event or Problem · 1
ON AN UNREPORTED DATE, THE PATIENT EXPERIENCED A BREAK IN ASEPTIC TECHNIQUE AND EXPERIENCED PERITONITIS, NOT REQUIRING HOSPITALIZATION. THE PATIENT WAS TREATED WITH INJECTIONS OF VANCOMYCIN-INTRAPERITONEALLY (IP) 1GM, AND REFLIN IP 1 GM (FREQUENCIES WERE NOT REPORTED). THE PATIENT IS RECOVERING FROM THE PERITONITIS. NO ADDITIONAL INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 249982 | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention | DIANEAL |