FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC PLUS BED

MDR report key: 3150373 · Received May 31, 2013

Report

Report Number
1824206-2013-02928
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 4, 2013
Report Date
May 4, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE SERVICE TECH FOUND THAT THERE WAS A PLASTIC CAP IN THE SIDE RAIL LATCH MECHANISM. THE PLASTIC CAP WAS REMOVED FROM THE SIDE RAIL LATCH MECHANISM TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THAT THE SIDE RAIL WOULD NOT RAISE AND LATCH. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
240439 TOTALCARE BARIATRIC PLUS BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1