AUTOPULSE NIMH BATTERY
Report
- Report Number
- 3003793491-2013-00631
- Event Type
- Malfunction
- Date Received
- May 30, 2013
- Date of Event
- May 7, 2013
- Report Date
- May 7, 2013
- Manufacturer
- ZOLL CIRCULATION, INC
- Product Code
- DRM
- PMA / PMN Number
- K112998
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE AUTOPULSE BATTERY WITH SN (B)(4) WAS RETURNED TO ZOLL CIRCULATION ON 05/14/2013 FOR INVESTIGATION. VISUAL INSPECTION REVEALED NO ANOMALIES. CUSTOMER'S REPORTED PROBLEM WAS CONFIRMED. THE RETURNED BATTERY FAILED TESTING. BASE ON THE EVALUATION RESULTS, A PROBABLE ROOT CAUSE FOR CUSTOMER'S REPORTED COMPLAINT MAY BE DUE TO A BATTERY PROBLEM; HOWEVER, A DEFINITIVE ROOT CAUSE FOR THE BATTERY PROBLEM COULD NOT BE DETERMINED. PLEASE SEE THE FOLLOWING RELATED MFR REPORT NUMBERS: 3003793491-2013-00630 FOR AUTO PULSE RESUSCITATION SYS MODEL 100 WITH SN (B)(4), 3003793491-2013-00632 FOR AUTO PULSE NIMH BATTERY WITH SN (B)(4) AND 3003793491-2013-00633 FOR AUTO PULSE NIMH BATTERY WITH SN (B)(4).
IT WAS REPORTED THAT THE AUTOPULSE PLATFORM WAS USED ON A (B)(6) FEMALE IN CARDIAC ARREST. PT SIZE WAS AVERAGE. THE PLATFORM COMPRESSED FOR 2 MINUTES AND WAS STOPPED BY THE OPERATORS. WHEN THEY TRIED TO RESTART, THE LIFEBAND WOULD RESIZE, PERFORM 5 COMPRESSIONS AND THEN STOP. THE DISPLAY DARKENED AND PROMPTED "REPLACE BATTERY". THE USERS ATTEMPTED TO CONTINUE A COUPLE MORE TIMES BUT REVERTED TO MANUAL CPR. NO ADVERSE PT SEQUELAE WAS REPORTED. THE CUSTOMER UTILIZES A 3-BATTERY ROTATION AND STATED THAT THEY TEST CYCLE THEIR BATTERIES MONTHLY BUT ROTATE BATTERIES EVERY 5 DAYS. THE CUSTOMER WAS INFORMED THAT THE CURRENT RECOMMENDATIONS ARE TO ROTATE THE BATTERIES DAILY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 238296 | AUTOPULSE NIMH BATTERY | AUTOPULSE NIMH BATTERY | DRM | ZOLL CIRCULATION, INC | 8700-0702 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | MANUAL CPR |