FDA Adverse Event Malfunction Summary report: N

AUTOPULSE NIMH BATTERY

MDR report key: 3150319 · Received May 30, 2013

Report

Report Number
3003793491-2013-00631
Event Type
Malfunction
Date Received
May 30, 2013
Date of Event
May 7, 2013
Report Date
May 7, 2013
Manufacturer
ZOLL CIRCULATION, INC
Product Code
DRM
PMA / PMN Number
K112998
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE AUTOPULSE BATTERY WITH SN (B)(4) WAS RETURNED TO ZOLL CIRCULATION ON 05/14/2013 FOR INVESTIGATION. VISUAL INSPECTION REVEALED NO ANOMALIES. CUSTOMER'S REPORTED PROBLEM WAS CONFIRMED. THE RETURNED BATTERY FAILED TESTING. BASE ON THE EVALUATION RESULTS, A PROBABLE ROOT CAUSE FOR CUSTOMER'S REPORTED COMPLAINT MAY BE DUE TO A BATTERY PROBLEM; HOWEVER, A DEFINITIVE ROOT CAUSE FOR THE BATTERY PROBLEM COULD NOT BE DETERMINED. PLEASE SEE THE FOLLOWING RELATED MFR REPORT NUMBERS: 3003793491-2013-00630 FOR AUTO PULSE RESUSCITATION SYS MODEL 100 WITH SN (B)(4), 3003793491-2013-00632 FOR AUTO PULSE NIMH BATTERY WITH SN (B)(4) AND 3003793491-2013-00633 FOR AUTO PULSE NIMH BATTERY WITH SN (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE AUTOPULSE PLATFORM WAS USED ON A (B)(6) FEMALE IN CARDIAC ARREST. PT SIZE WAS AVERAGE. THE PLATFORM COMPRESSED FOR 2 MINUTES AND WAS STOPPED BY THE OPERATORS. WHEN THEY TRIED TO RESTART, THE LIFEBAND WOULD RESIZE, PERFORM 5 COMPRESSIONS AND THEN STOP. THE DISPLAY DARKENED AND PROMPTED "REPLACE BATTERY". THE USERS ATTEMPTED TO CONTINUE A COUPLE MORE TIMES BUT REVERTED TO MANUAL CPR. NO ADVERSE PT SEQUELAE WAS REPORTED. THE CUSTOMER UTILIZES A 3-BATTERY ROTATION AND STATED THAT THEY TEST CYCLE THEIR BATTERIES MONTHLY BUT ROTATE BATTERIES EVERY 5 DAYS. THE CUSTOMER WAS INFORMED THAT THE CURRENT RECOMMENDATIONS ARE TO ROTATE THE BATTERIES DAILY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
238296 AUTOPULSE NIMH BATTERY AUTOPULSE NIMH BATTERY DRM ZOLL CIRCULATION, INC 8700-0702 NA

Patients

Seq Age Sex Outcome Treatment
1 78 YR MANUAL CPR