OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2013-00510
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- April 29, 2013
- Report Date
- April 29, 2013
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE KINKED CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT QUALIFICATION RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER," AND "IF YOUR READING IS ABOVE 500 MG/DL, THE PDM DISPLAYS 'HIGH CHECK FOR KETONES!' THIS INDICATES SEVERE HYPERGLYCEMIA (HIGH BLOOD GLUCOSE). IF YOU GET A 'HIGH CHECK FOR KETONES!' READING AND FEEL SYMPTOMS SUCH AS FATIGUE, THIRST, EXCESS URINATION, OR BLURRY VISION, FOLLOW YOUR HEALTHCARE PROVIDER'S RECOMMENDATION TO TREAT HYPERGLYCEMIA. IF YOU GET A 'HIGH CHECK FOR KETONES!' READING, BUT HAVE NO SYMPTOMS OF HIGH BLOOD GLUCOSE, THEN RETEST WITH A NEW TEST STRIP ON YOUR FINGERS. IF YOU STILL GET A 'HIGH CHECK FOR KETONES!' READING, PERFORM A CONTROL SOLUTION TEST TO ENSURE YOUR SYSTEM IS WORKING PROPERLY. IF THE SYSTEM IS WORKING PROPERLY, FOLLOW YOUR HEALTHCARE PROVIDER'S RECOMMENDATION TO TREAT HYPERGLYCEMIA." IT ADVISES, "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE NOT PRESENT, FOLLOW YOUR HEALTHCARE PROVIDER'S GUIDELINES. IT KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE AGAIN AFTER 2 HOURS. IF BLOOD GLUCOSE LEVELS HAVE NOT DECREASED, TAKE A SECOND BOLUS BY INJECTION, USING A STERILE SYRINGE. IF BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER 2 HOURS (A TOTAL OF 4 HOURS), REPLACE THE POD."
THE CUSTOMER REPORTED THAT SHE ACTIVATED A POD ON (B)(6) AT 11:26 PM. SHE PROVIDED THE FOLLOWING HISTORY FOR ON (B)(6): TIME: 12:16 AM, BLOOD GLUCOSE: 500 MG/DL, BOLUS: 5.20U; TIME: 2:11 AM, BLOOD GLUCOSE: 367 MG/DL; TIME: 6:05 AM, BLOOD GLUCOSE: 435 MG/DL, BOLUS: 6.70U. AT 6:51 AM, WITH A RESULT OF "HIGH" (>500 MG/DL), SHE DEACTIVATED THE POD. SHE SAID THAT THE CANNULA WAS KINKED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233743 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L31072 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |