FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 3150267 · Received May 29, 2013

Report

Report Number
3004464228-2013-00509
Event Type
Malfunction
Date Received
May 29, 2013
Date of Event
April 29, 2013
Report Date
April 29, 2013
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. WE ARE UNABLE TO CONFIRM THE BENT/KINKED CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. LOT QUALIFICATION RECORDS WERE REVIEWED, AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD USER GUIDE WARNS "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER," AND "IF YOUR READING IS ABOVE 500 MG/DL, THE PDM DISPLAYS 'HIGH CHECK FOR KETONES!' THIS INDICATES SEVERE HYPERGLYCEMIA (HIGH BLOOD GLUCOSE). IF YOU GET A 'HIGH CHECK FOR KETONES!' READING AND FEEL SYMPTOMS SUCH AS FATIGUE, THIRST, EXCESS URINATION, OR BLURRY VISION, FOLLOW YOUR HEALTHCARE PROVIDER'S RECOMMENDATION TO TREAT HYPERGLYCEMIA. IF YOU GET A 'HIGH CHECK FOR KETONES!' READING, BUT HAVE NO SYMPTOMS OF HIGH BLOOD GLUCOSE, THEN RETEST WITH A NEW TEST STRIP ON YOUR FINGERS. IF YOU STILL GET A 'HIGH CHECK FOR KETONES!' READING, PERFORM A CONTROL SOLUTION TEST TO ENSURE YOUR SYSTEM IS WORKING PROPERLY. IF THE SYSTEM IS WORKING PROPERLY, FOLLOW YOUR HEALTHCARE PROVIDER'S RECOMMENDATION TO TREAT HYPERGLYCEMIA." IT ADVISES, "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE NOT PRESENT, FOLLOW YOUR HEALTHCARE PROVIDER'S GUIDELINES. IT KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE AGAIN AFTER 2 HOURS. IF BLOOD GLUCOSE LEVELS HAVE NOT DECREASED, TAKE A SECOND BOLUS BY INJECTION, USING A STERILE SYRINGE. IF BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER 2 HOURS (A TOTAL OF 4 HOURS), REPLACE THE POD."

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT SHE ACTIVATED A POD ON (B)(6) AT 7:17 PM, WITH A BLOOD GLUCOSE RESULT OF 283 MG/DL. SHE THEN CONSUMED 35 GRAMS OF CARBOHYDRATE AND GAVE HERSELF A 6.45 UNIT BOLUS. SHE REPORTED THE FOLLOWING HISTORY FOR (B)(6). AT 11:03 PM, HER RESULT WAS "HIGH" (>500 MG/DL), AND SHE GAVE HERSELF A MANUAL INJECTION OF 10U OF NOVOLOG. AT 11:21 PM, SHE DEACTIVATED THE POD. SHE SAID THAT THE CANNULA WAS BENT/KINKED IN AN L SHAPE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233605 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L31072

Patients

Seq Age Sex Outcome Treatment
1 35 YR