LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01396
- Event Type
- Malfunction
- Date Received
- May 29, 2013
- Date of Event
- April 18, 2013
- Report Date
- May 20, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER ON) HAS BEEN CONFIRMED. AS RECEIVED, THE MONITOR WAS RESETTING. THE CAUSE OF THE RESETS WAS AN INTERMITTENT CONNECTION AT BGA COMPONENT U500 (DSP). THE INTERMITTENT CONNECTION IS LIKELY DUE TO A FRACTURED BGA SOLDER CONNECTION INDUCED BY MECHANICAL STRESS. THE EXACT SOURCE OF THE MECHANICAL STRESS HAS NOT BEEN POSITIVELY IDENTIFIED BUT THE FRACTURED CONNECTION MAY HAVE BEEN ACCELERATED THROUGH ROUGH HANDLING, AS THE MONITOR CASE WAS RETURNED WITH SIGNS OF DAMAGE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE ((B)(4)) TO REDUCE THE PCA STRAIN WAS APPROVED BY FDA ON (B)(4) 2012. IMPLEMENTATION BEGAN ON (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULD NOT POWER ON PROPERLY. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 233670 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |