FDA Adverse Event Malfunction Summary report: N

INRATIO ASSY, BOX OF STRIPS, MULTILANGUAGE

MDR report key: 3150125 · Received June 3, 2013

Report

Report Number
2027969-2013-00443
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 24, 2013
Report Date
June 3, 2013
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K092987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IMPROPER TECHNIQUES WERE IDENTIFIED BY (B)(4). IMPROPER TECHNIQUES MAY LEAD TO INACCURATE RESULTS. INRATIO PRECISION DATA PROVIDED BY END-USER: DATE: (B)(6) 2013, 1ST INR = 5.1, 2ND INR = 2.1, MEAN = 3.6, SD = 2.12, %CV = 58.93. INRATIO METER RESULTS FAILED THE CRITERIA FOR PRECISION WITH %CV GREATER THAN (B)(4). IN-HOUSE THERAPEUTIC DONOR TESTING WAS PERFORMED ON REPORTED LOT # 305771 ON (B)(4) 2013. RESULTS AS FOLLOWS: (B)(4). NO FURTHER INVESTIGATION WAS REQUIRED. ANALYSIS OF THE CLIENT'S DATA FROM REPEAT INRATIO TESTING REVEALED THAT THE TEST RESULT COMPARISON DID NOT MEET PRECISION CRITERIA. NO PRODUCT WAS EXPECTED TO BE RETURNED SO RETAIN PRODUCTS WERE USED OF INVESTIGATION TESTING. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND PRECISION CRITERIA. (B)(4). NO CORRECTIVE ACTION AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT IMPRECISION RESULTS. RESULTS AS FOLLOWS: DATE: (B)(6) 2013, IRATIO: 5.1, REPEAT: 2.1. TIME BETWEEN TESTING WAS REPORTED AS 3 MINUTES. PT'S THERAPEUTIC RANGE IS 2.0 - 3.0. IT WAS REPORTED THE PT SELF TESTER (PST) WAS MILKING THEIR FINGER AND TOUCHING THEIR FINGER TO THE STRIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242895 INRATIO ASSY, BOX OF STRIPS, MULTILANGUAGE PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100071 305771

Patients

Seq Age Sex Outcome Treatment
1 COUMADIN