FDA Adverse Event Injury Summary report: N

DAVOL SILICONE ROUND SINGLE DRAIN

MDR report key: 3150091 · Received May 31, 2013

Report

Report Number
1018233-2013-02315
Event Type
Injury
Date Received
May 31, 2013
Report Date
May 3, 2013
Manufacturer
PRODUCTOS PARA EL CUIDADO DE LA SALUD
Product Code
GBX
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS INCONCLUSIVE AS NO SAMPLE WAS RETURNED FOR EVAL. REVIEW OF THE DEVICE HISTORY RECORD FOR THE INVOLVED LOT NUMBER DID NOT FIND ANY DISCREPANCY THAT COULD HAVE CONTRIBUTED TO THE ADVERSE EVENT OF DRAIN BREAKAGE. THE INSTRUCTIONS FOR USE PROVIDE THE FOLLOWING: ADD'L PERFORATIONS SHOULD NOT BE MADE IN THE DRAINS. AVOID SUTURING THROUGH DRAINS. DRAINS SHOULD LIE FLAT AND IN LINE WITH THE SKIN EXIT AREAS. PARTICULAR CARE SHOULD BE TAKEN TO AVOID ANY OBSTACLES TO THE DRAIN EXIT PATH. DRAINS SHOULD BE CHECKED FOR FREE MOTION DURING CLOSURE TO MINIMIZE THE POSSIBILITY OF BREAKAGE. DRAIN REMOVAL SHOULD BE DONE GENTLY BY HAND. THEY SHOULD NOT BE HANDLED WITH POINTED, TOOTHED OR SHARP INSTRUMENTS WHICH COULD CAUSE CUTS OR NICKS AND LEAD TO SUBSEQUENT STRUCTURAL FAILURE OF THE DRAIN. SURGICAL REMOVAL MAY BE NECESSARY IF DRAIN IS DIFFICULT TO REMOVE OR BREAKS. (B)(4).

Description of Event or Problem · 1

(B)(4). THE PATIENT HAD A SURGICAL PROCEDURE WHICH INVOLVED PLACEMENT OF A DRAIN IN THEIR NECK. IT WAS REPORTED THAT UPON REMOVAL OF THE DRAIN, IT BROKE AND APPROX 2 INCHES WAS LEFT IN THE PATIENT'S NECK. THE DRAIN HAD BEEN SUTURED IN PLACE WITH LIGHT SILK TO MAINTAIN POSITION FROM ONE SIDE OF THE NECK TO THE OTHER. IT WAS FURTHER REPORTED THAT A SECOND PROCEDURE WAS PERFORMED TO REMOVE THE DRAIN WHICH INVOLVED A RE-INCISION IN THE CHIN TO EXTRACT THE DRAIN PIECES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
240174 DAVOL SILICONE ROUND SINGLE DRAIN GBX PRODUCTOS PARA EL CUIDADO DE LA SALUD NA UNK

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention