FDA Adverse Event Malfunction Summary report: N

ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT

MDR report key: 3150022 · Received May 20, 2013

Report

Report Number
1419322-2013-00014
Event Type
Malfunction
Date Received
May 20, 2013
Report Date
April 22, 2013
Manufacturer
DENTSPLY PROFESSIONAL
Product Code
EKX
PMA / PMN Number
K031145
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THOUGH (B)(4) MEDICAL/ SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/ SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE 1:5 ESTYLUS ATTACHMENT REACHED 97.6 DEGREES CELSIUS AFTER ONLY 27 SECONDS OF FREE RUN TESTING, EXCEEDING ISO AND ES1104 STANDARDS FOR MAXIMUM ALLOWABLE TEMPERATURE DUE TO BUR END BEARING FAILURE, GEAR WEAR AND EXCESSIVE DEBRIS.

Description of Event or Problem · 1

IN THIS EVENT A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.