FDA Adverse Event
Malfunction
Summary report: N
ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT
MDR report key: 3150022
·
Received May 20, 2013
Report
- Report Number
- 1419322-2013-00014
- Event Type
- Malfunction
- Date Received
- May 20, 2013
- Report Date
- April 22, 2013
- Manufacturer
- DENTSPLY PROFESSIONAL
- Product Code
- EKX
- PMA / PMN Number
- K031145
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- DENTIST
Narratives
Additional Manufacturer Narrative · 1
THOUGH (B)(4) MEDICAL/ SURGICAL INTERVENTION WAS REQUIRED TO PRECLUDE A SERIOUS INJURY IN THIS EVENT, THERE HAVE BEEN PREVIOUSLY REPORTED EVENTS INVOLVING THIS DEVICE THAT RESULTED IN THE NEED FOR MEDICAL/ SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION. THEREFORE, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE 1:5 ESTYLUS ATTACHMENT REACHED 97.6 DEGREES CELSIUS AFTER ONLY 27 SECONDS OF FREE RUN TESTING, EXCEEDING ISO AND ES1104 STANDARDS FOR MAXIMUM ALLOWABLE TEMPERATURE DUE TO BUR END BEARING FAILURE, GEAR WEAR AND EXCESSIVE DEBRIS.
Description of Event or Problem · 1
IN THIS EVENT A DOCTOR REPORTED THAT AN ESTYLUS 1:5 HANDPIECE OVERHEATED. THERE WAS NO INJURY OR INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 222914 | ESTYLUS 1:5 CONTRA ANGLE ATTACHMENT | EKX | DENTSPLY PROFESSIONAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|