MONOPOLAR CURVED SHEARS INSTRUMENT

MDR report key: 3150014 · Received June 6, 2013

Report

Report Number
2955842-2013-02014
Event Type
Malfunction
Date Received
June 6, 2013
Date of Event
May 6, 2013
Report Date
May 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT. THE INSTRUMENT WAS CHECKED FOR RECOGNITION ON AN IN-HOUSE SYSTEM AND THE INSTRUMENT WAS SUCCESSFULLY RECOGNIZED ON MULTIPLE ATTEMPTS. THE POGO PINS WERE NOT STUCK OR CONTAMINATED. ENGINEERING WAS UNABLE TO REPLICATE RECOGNITION ISSUE. THE INSTRUMENT WAS DRIVEN AND MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ADDITIONAL OBSERVATION THAT WAS NOT REPORTED BY THE CUSTOMER WERE MICRO-CRACKS FOUND AT THE DISTAL END OF THE TUBE. THE MICRO-CRACKS RUN IN THE AXIAL DIRECTION AND MEASURES APPROXIMATELY 0.334. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT, HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESE IS CONFINED TO 2CM OF THE DISTAL END OF THE INSTRUMENT SHAFT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI PROCEDURE THE MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT WAS NOT RECOGNIZED BY THE SYSTEM. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
250125 MONOPOLAR CURVED SHEARS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420179-10 M11130208 147

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES