FDA Adverse Event
Malfunction
Summary report: N
CXDI CCS NE VER.1.40.2
MDR report key: 3150002
·
Received May 17, 2013
Report
- Report Number
- 1000181430-2013-00090
- Event Type
- Malfunction
- Date Received
- May 17, 2013
- Date of Event
- April 18, 2013
- Report Date
- April 19, 2013
- Manufacturer
- CANON, INC.
- Product Code
- MQB
- PMA / PMN Number
- K102012
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THAT WAS A SAME ERROR WHICH WE HAVE EXPERIENCED AND ANALYZED IN THE PAST. THIS ERROR CAUSED BY THE TIMING AT WHICH CONNECTING OR DISCONNECTING WITH MEMORY IS PROCESSED WHEN EXPOSURES ARE TAKEN AND THE TIMING AT WHICH THE PREVIEW IMAGE ARE STORED. IT WAS RESULTED IN FAILURE TO STORE RAW IMAGES. THE PROBLEM HAS BEEN CORRECTED IN A SUBSEQUENT SOFTWARE UPGRADE (VERSION 2.02). (B)(4).
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEY RECEIVED AN ERROR E050600002 IMAGE FILE READING FAILED DURING OUTPUT IMAGE GENERATION. A TOTAL OF 5 IMAGES WERE LOST. THE DATA WAS AMONG 3 PTS; 1 IMAGE LOST ON ONE PT, 2 IMAGES LOST ON THE OTHER 2 PTS. THE PTS HAD TO HAVE THE IMAGES REPEATED. THE IMAGE WAS RETAKEN, RESULTING IN ADDITIONAL RADIATION EXPOSURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 219555 | CXDI CCS NE VER.1.40.2 | MQB | CANON, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | CXDI-70C WIRELESS |