FDA Adverse Event Malfunction Summary report: N

MEDIHONEY GEL

MDR report key: 3149097 · Received May 31, 2013

Report

Report Number
MW5030414
Event Type
Malfunction
Date Received
May 31, 2013
Report Date
May 20, 2013
Manufacturer
DERMA SCIENCES
Product Code
FRO
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

FOUR OUT OF 4 COMMERCIAL LOTS OF MEDIHONEY GEL WERE TESTED FOR THE PRESENCE OF ENDOTOXIN AND FOUND TO HAVE LEVELS IN EXCESS OF THE USP LIMIT OF 0.5 EU/ML FOR DEVICES THAT DIRECTLY OR INDIRECTLY CONTACT THE CARDIOVASCULAR OR LYMPHATIC SYSTEM. THESE LOTS WERE SUBMITTED TO AN INDEPENDENT LABORATORY FOR CONFIRMATORY TESTING. RESULTS CONFIRM LEVELS THAT ARE 1,000-2,000 TIMES THE UPPER ACCEPTANCE LIMIT. WHEREAS THE PRODUCT IS USED IN DIRECT REPEATED APPLICATION TO OPEN WOUNDS, INCLUDING IN NEONATES AND THE ELDERLY, THESE HIGHLY ELEVATED ENDOTOXIN LEVELS APPEAR HAZARDOUS. EXACERBATED WOUND INFLAMMATION, DELAYED HEALING, AND IATROGENIC SYSTEMIC INFLAMMATORY RESPONSES MAY RESULT FROM USE OF AFFECTED LOTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
240354 MEDIHONEY GEL MEDICAL GRADE HONEY DRESSING FRO DERMA SCIENCES 1133
240355 MEDIHONEY GEL MEDICAL GRADE HONEY DRESSING FRO DERMA SCIENCES 1213
240356 MEDIHONEY GEL MEDICAL GRADE HONEY DRESSING FRO DERMA SCIENCES 1225
240478 MEDIHONEY GEL MEDICAL GRADE HONEY DRESSING FRO DERMA SCIENCES 1219

Patients

Seq Age Sex Outcome Treatment
1