FDA Adverse Event Injury Summary report: N

UNISEN STAR TRAK

MDR report key: 314777 · Received February 2, 2001

Report

Report Number
314777
Event Type
Injury
Date Received
February 2, 2001
Date of Event
December 12, 2000
Report Date
January 16, 2001
Manufacturer
UNISEN, INC
Product Code
IOL
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
FL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

THE PT WAS PARTICIPATING IN AN OUTPATIENT PULMONARY REHABILITATION PROGRAM. WHILE WALKING ON THE TREADMILL THE PT ATTEMPTED TO INCREASE THE SPEED FROM 1.6 TO 1.8 MPH. APPARENTLY THE PT ATTEMPTED TO DEPRESS THE #1 ON THE DISPLAY PANEL AND THEN THE #8. INADVERTENTLY NO CONTACT WAS MADE WITH THE NUMBER 1 AND THE SPEED BEGAN RAPIDLY INCREASING TO REACH 8 MPH. THE PT WAS THROWN BACKWARDS OFF THE TREADMILL AND SUSTAINED IN ADDITION TO MULTIPLE ABRASIONS AND BRUISES A LEFT SHOULDER DISLOCATION AND FRACTURE TO THE HUMERAL HEAD. SURGICAL INTERVENTION WAS REQUIRED. THE PT ALSO REQUIRED REHABILITATIVE AND HOME HEALTH SERVICES. ALTHOUGH UNABLE TO DETERMINE IF RELATED TO THE EVENT, THE PT EXPIRED 19 POST EVENT WHILE AT HOME. FOLLOWING THE EVENT, THE FACILITY DISCUSSED SITUATION WITH THE VENDOR AND MODIFIED DISPLAY/CONTROL PANELS. GIVEN THE AGE OF FACILITY'S PT POPULATION, IT IS BELIEVED THAT THE PT INJURY MAY BE A RESULT OF PRODUCT DESIGN RATHER THAN A DEFECT IN THE EQUIPMENT WHICH EASILY LED TO USER ERROR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4085 UNISEN STAR TRAK TREADMILL IOL UNISEN, INC STAR TRAK 2000 NA

Patients

Seq Age Sex Outcome Treatment
1 81 YR Hospitalization| R