FDA Adverse Event Malfunction Summary report: N

LINEAR SERIES COLLIMATORS

MDR report key: 314486 · Received January 29, 2001

Report

Report Number
1423380-2001-00001
Event Type
Malfunction
Date Received
January 29, 2001
Report Date
January 25, 2001
Manufacturer
PROGENY, INC.
Product Code
KPW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

CERTAIN BATCHES OF COLLIMATORS BUILT IN SEP-OCT 1999 HAVE INCORRECT (SHORTER) SCREWS MOUNTING THE COLLIMATOR HOUSING TO THE TUBE MOUNTING PLATE. THE SHORTER SCREWS OVER TIME CAN LOOSEN, ALLOWING THE COLLIMATOR HOUSING TO SAG RELATIVE TO THE MOUNTING PLATE. IF THE "LOOSENESS" OF THE HOUSING WAS NOT NOTICED BY STAFF, THE HOUSING COULD EVENTUALLY SEPARATE FROM THE MOUNTING PLATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3433 LINEAR SERIES COLLIMATORS COLLIMATOR KPW PROGENY, INC. LINEAR MC150, LINEAR II, LINEAR III, LINEAR IV *

Patients

Seq Age Sex Outcome Treatment
1 NA