FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3143736 · Received June 3, 2013

Report

Report Number
2210968-2013-06437
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
December 14, 2012
Report Date
May 6, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INFORMATION CONTAINED IN THIS FILE HAS BEEN DETERMINED TO BE A DUPLICATE OF THE EVENT REPORTED IN MEDWATCH 2210968-2013-10007. THEREFORE, THIS MEDWATCH 2210968-2013-06437 WILL BE VOIDED. ALL INFORMATION REGARDING THIS EVENT WILL RESIDE IN MEDWATCH 2210968-2013-10007.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE DEVICE WAS RECEIVED IN A CONDITION WHICH DOES NOT MEET SPECIFICATION. THE BLADE FAILED TO ROTATE DURING THE EVALUATION; IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-00753. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2012. DURING THE PROCEDURE THE DEVICE SEIZED UP DURING MORCELLATING. THE PROCEDURE WAS COMPLETED WITH NO ADVERSE PATIENT CONSEQUENCES.