MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2013-01967
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- March 18, 2013
- Report Date
- May 6, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE CONDUCTOR WIRE WAS BROKEN AT THE YAW PULLEY EXIT. THE WIRE WAS DETACHED FROM THE CONNECTION AT THE GRIP. THE INSTRUMENT FAILED ELECTRICAL CONTINUITY TESTING. THE YAW PULLEY SHOWED NO SIGNS OF ARCING. THERE WERE SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. ENGINEERING ALSO FOUND THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS SHOWING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE SHORT IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE MARYLAND BIPOLAR FORCEPS INSTRUMENT WIRES BROKE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243977 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10130115 249 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |