MONOPOLAR CURVED SCISSORS INSTRUMENT
Report
- Report Number
- 2955842-2013-01965
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- April 23, 2013
- Report Date
- May 6, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050005
- Removal / Correction Number
- 2955842-051613-005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- NURSE
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE INSTRUMENT WAS PLACED ON SYSTEM FOR LATEX CUT TEST AND THE SCISSORS DID NOT CUT CLEANLY THROUGH .006 LATEX. THE LATEX WAS SNAGGED AT THE SCISSOR TIPS. THE BLADE EDGES WERE UNDAMAGED, BUT EDGES EXHIBITED WEAR AT THE TIPS, NEGATIVELY AFFECTING CUT PERFORMANCE. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. DEEP SCRATCHES NEAR THE DISTAL TIP OF THE INSTRUMENT APPROXIMATELY 0.8 FROM THE BASE OF THE TUBE EXTENSION. ADDITIONAL OBSERVATION THAT WAS NOT REPORTED BY THE CUSTOMER WAS MICRO-CRACKS FOUND AT THE DISTAL END OF THE TUBE. THE MICRO-CRACKS RUN IN THE AXIAL DIRECTION. THESE TYPES OF MICRO-CRACKS WILL NOT LEAD TO MECHANICAL FAILURE OF THE INSTRUMENT, HOWEVER, THERE IS A POTENTIAL FOR INSULATION FAILURE AFTER REPROCESSING, RESULTING IN A PATHWAY FOR ELECTROSURGICAL ENERGY TO LEAK TO TISSUE AND POTENTIALLY CAUSE UNINTENDED INJURIES. THE LOCATION OF THESE IS CONFINED TO 2CM OF THE DISTAL END OF THE INSTRUMENT SHAFT. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGE TO THE INSTRUMENT'S MAIN TUBE, WHICH WAS FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI SI GASTRIC BYPASS PROCEDURE THE MONOPOLAR CURVED SCISSORS INSTRUMENT WAS NOTED TO BE DULL. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245152 | MONOPOLAR CURVED SCISSORS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420179-10 | M10121126 470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |