3.5MM 90-S SERFAS ENERGY SUCTION PROBE
Report
- Report Number
- 0002936485-2013-00243
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 9, 2013
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- GEI
- PMA / PMN Number
- K071859
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE REPORTED FAILURE MODE WAS CONFIRMED. THE MISSING FLOWER SHAPED ELECTRODE PORTION WAS RETURNED FOR EVALUATION. THE REST OF THE ELECTRODE WAS VISIBLE INSIDE THE CERAMIC TIP. TISSUE RESIDUE WAS OBSERVED ON THE ELECTRODE AND THE CERAMIC. SCRATCH WEAR MARKS WERE OBSERVED ON ALL OF THE CERAMIC AND LUMEN. NO VISUAL WEAR MARKS WERE OBSERVED ON THE PROBE¿S INSULATION. REVIEW OF THE DEVICE HISTORY RECORD OF THE REPORTED LOT NUMBER WAS REVIEWED. THE REVIEW DISCLOSED NO DISCREPANCIES THAT COULD CONTRIBUTE TO THIS CONDITION. THE MOST PROBABLE ROOT CAUSE IS DESIGN RELATED WITH A USER RELATED (MISUSE) POSSIBLE CONTRIBUTOR FACTOR. IN SUM, THE PRODUCT WAS RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE WAS CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
IT WAS REPORTED THAT DURING THE PROCEDURE THE WAND OF THE UNIT BROKE INSIDE OF THE PATIENT. A FIVE MINUTE DELAY WAS REPORTED AND ALLEGEDLY THERE WAS HARM TO THE PATIENT WITH NO ADVERSE CONSEQUENCE. IT WAS FURTHER REPORTED THAT THE TIP WAS RETRIEVED FROM THE PATIENT.
IT WAS REPORTED THAT DURING THE PROCEDURE THE WAND OF THE UNIT BROKE INSIDE OF THE PATIENT. A FIVE MINUTE DELAY WAS REPORTED AND ALLEGEDLY THERE WAS HARM TO THE PATIENT WITH NO ADVERSE CONSEQUENCE. IT WAS FURTHER REPORTED THAT THE TIP WAS RETRIEVED FROM THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245071 | 3.5MM 90-S SERFAS ENERGY SUCTION PROBE | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | STRYKER ENDOSCOPY-SAN JOSE | 12205AE2 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |