ADVIA CENTAUR XP TSH3-ULTRA (TSH3-UL) ASSAY
Report
- Report Number
- 1219913-2013-00118
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- October 24, 2012
- Report Date
- May 7, 2013
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- JLW
- PMA / PMN Number
- K083844
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
SIEMENS INITIALLY ASSESSED THIS EVENT AS NO POTENTIAL FOR INJURY; THEREFORE, NOT REPORTABLE. AFTER COMPLETION OF INTERNAL INVESTIGATIONS, A ROOT CAUSE FOR THE DISCORDANT RESULTS WAS DETERMINED, LEADING TO A NEW AWARENESS DATE FOR THIS EVENT OF (B)(6) 2013. INVESTIGATIONS CONCLUDED THAT A RARE VARIANT OF TSH, IDENTIFIED IN A SMALL CLUSTER OF PATIENTS, IS NOT DETECTED BY SIEMENS ADVIA CENTAUR SYSTEMS TSH3-ULTRA ASSAY. INVESTIGATION OF UNEXPLAINED DISCORDANT ADVIA CENTAUR TSH3-ULTRA RESULTS IDENTIFIED SAMPLES IN WHICH THE MONOCLONAL ANTIBODY USED IN THE REAGENT FAILED TO DETECT THE TSH MOLECULE. THESE INDIVIDUALS MAY HAVE A PREVIOUSLY UNRECOGNIZED, FUNCTIONALLY NORMAL TSH VARIANT. (B)(4). AS A RESULT, AN IMPORTANT PRODUCT SAFETY INFORMATION BULLETIN (B)(4) WAS SENT TO CUSTOMERS IN THE UNITED STATES AND AN URGENT FIELD SAFETY NOTICE (B)(4) WAS SENT TO CUSTOMERS OUTSIDE OF THE UNITED STATES IN (B)(4) 2013, TO INFORM THEM OF THIS ISSUE. THESE CUSTOMER COMMUNICATIONS ALSO INCLUDED A REMINDER THAT, FOR DIAGNOSTIC PURPOSES, THE RESULTS OBTAINED FROM THESE ASSAYS SHOULD ALWAYS BE USED IN COMBINATION WITH THE CLINICAL EXAMINATION, PATIENT MEDICAL HISTORY, AND OTHER FINDINGS.
FALSELY LOW ADVIA CENTAUR XP TSH3-ULTRA TEST RESULTS WERE OBTAINED BY THE CUSTOMER AND THE RESULTS WERE CONSIDERED DISCORDANT WHEN COMPARED TO ALTERNATE TEST METHOD RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT ADVIA CENTAUR XP TSH3-ULTRA RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243005 | ADVIA CENTAUR XP TSH3-ULTRA (TSH3-UL) ASSAY | TSH3-ULTRA IMMUNOASSAY | JLW | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | N/A | 258 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |