FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3143512 · Received June 3, 2013

Report

Report Number
2955842-2013-01959
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 14, 2013
Report Date
May 16, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. RECEIVED THE INSTRUMENT INVOLVED WITH THE COMPLAINT. ENGINEERING CONFIRMED THAT THE PITCH UP CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP IS STILL INSTALLED IN CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. OTHER CABLES AT THE INSTRUMENT'S WRIST WERE NOT DAMAGED. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGEON REPORTEDLY NOTED THAT THE PROGRASP FORCEPS INSTRUMENT WAS NOT RESPONDING. THE INSTRUMENT WAS REMOVED FROM THE PATIENT AND THE USER FACILITY ALLEGEDLY FOUND BROKEN CABLES ON THE INSTRUMENT. NOTHING REPORTEDLY FELL INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
243303 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M10130318 526

Patients

Seq Age Sex Outcome Treatment
1 69 YR DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES