FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR

MDR report key: 3143405 · Received June 3, 2013

Report

Report Number
3015876-2013-00466
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 7, 2013
Report Date
May 7, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
MKJ
PMA / PMN Number
K103567
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT PROBLEM. PHYSIO RE-SEATED ALL OF THE CONNECTIONS TO THE BATTERY PINS AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND THEN RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE INTERMITTENTLY FAILS TO POWER ON OR IT SPORADICALLY LOSES POWER WHEN BATTERY ONE IS INSTALLED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.