FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
MDR report key: 3143405
·
Received June 3, 2013
Report
- Report Number
- 3015876-2013-00466
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 7, 2013
- Report Date
- May 7, 2013
- Manufacturer
- PHYSIO-CONTROL, INC
- Product Code
- MKJ
- PMA / PMN Number
- K103567
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4): PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED INTERMITTENT PROBLEM. PHYSIO RE-SEATED ALL OF THE CONNECTIONS TO THE BATTERY PINS AND OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS REPAIRED AND THEN RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE INTERMITTENTLY FAILS TO POWER ON OR IT SPORADICALLY LOSES POWER WHEN BATTERY ONE IS INSTALLED. THERE WAS NO PATIENT USE ASSOCIATED WITH THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244747 | LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR | DEFIBRILLATORS, AUTOMATIC, EXTERNAL | MKJ | PHYSIO-CONTROL, INC | 15 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |