FDA Adverse Event Malfunction Summary report: N

FOX SV PTA CATHETER

MDR report key: 3143356 · Received June 3, 2013

Report

Report Number
2024168-2013-03465
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 3, 2013
Report Date
May 9, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
LIT
PMA / PMN Number
K102705
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE RESISTANCE WITH THE GUIDE WIRE WAS UNABLE TO BE CONFIRMED HOWEVER THE DEFLATION DIFFICULTIES AND BALLOON REFOLD WERE ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LEFT SUPERFICIAL FEMORAL TO THE POPLITEAL ARTERIES. A 6 X 120 MM FOX SV WAS BEING USED TO POST DILATE A PREVIOUSLY PLACED UNKNOWN STENT. AFTER POST-DILATATION, THE BALLOON DEFLATED POORLY WITH A POOR REFOLD AND THERE WAS RESISTANCE REMOVING IT FROM A NON-ABBOTT GUIDE WIRE. THE PROCEDURE WAS COMPLETED AT THIS TIME AS THE STENT WAS WELL OPPOSED TO THE VESSEL WALL. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244541 FOX SV PTA CATHETER PERIPHERAL DILATATION CATHETER LIT AV-TEMECULA-CT 749568

Patients

Seq Age Sex Outcome Treatment
1 GUIDE CATH: 7F COOK