FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3143355
·
Received June 3, 2013
Report
- Report Number
- 1416980-2013-14134
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 20, 2013
- Report Date
- May 20, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE ROOT CAUSE OF THE BATTERY LOW ALARM WAS DETERMINED TO BE DAMAGE TO THE BATTERY. TO CORRECT THE CONDITION, THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A BATTERY LOW ALARM. NO ADDITIONAL INFORMATION IS AVAILABLE.