FDA Adverse Event Malfunction Summary report: N

PEDICSCR PANGEAPOLYAX Ø6 PREASSMBL L40 T

MDR report key: 3143279 · Received June 3, 2013

Report

Report Number
2530088-2013-00857
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 6, 2011
Report Date
May 9, 2011
Manufacturer
SYNTHES BRANDYWINE
Product Code
NKB
PMA / PMN Number
K103287
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND MANUFACTURING AND MATERIAL DOCUMENT SHOWS NO DEVIATION TO THE SPECIFICATION. NO PRODUCT FAULT COULD BE DETECTED. THE VISUAL INSPECTION REVEALED THE GOLD PART OF THE SCREW WAS DAMAGED. THE INVESTIGATION WAS NOT ABLE TO DETERMINE THE EXACT CAUSE WHICH LED TO THIS REPORTED PROBLEM. IT IS POSSIBLE THAT TOO MUCH MECHANICAL FORCE WAS APPLIED DURING THE REMOBILIZATION AND MAY HAVE CAUSED THE SEPARATION OF THE COMPONENTS. ANOTHER POSSIBLE REASON FOR SUCH AN OCCURRENCE COULD BE THAT SCREW HEAD BLOCKED AFTER INSERTION AND WITH THE ALIGNMENT TOOL ATTEMPTS TO ALIGN THE SCREW MAY HAVE STRIPPED THE SCREW HEAD. REVIEWING THE MANUFACTURING- AND MATERIAL DOCUMENT SHOWS NO DEVIATION TO THE SPECIFICATION. NO PRODUCT FAULT COULD BE DETECTED. THE INVESTIGATION DETERMINED THIS COMPLAINT TO BE INDETERMINATE. PLACEHOLDER.

Description of Event or Problem · 1

THE GOLD PART OF THE SCREW DAMAGED. DURING SURGERY AFTER ALIGNMENT WITH THE ALIGNMENT TOOL, THE DOCTOR NOTICED THAT THE GOLD PART OF THE SCREW WAS PROTRUDING. THE SCREW BROKE. THIS IS 1 OF 1 REPORT FOR THIS COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244404 PEDICSCR PANGEAPOLYAX Ø6 PREASSMBL L40 T NKB SYNTHES BRANDYWINE 6569495

Patients

Seq Age Sex Outcome Treatment
1