FDA Adverse Event Other Summary report: N

S3 DOUBLE HEAD PUMP

MDR report key: 3143234 · Received May 30, 2013

Report

Report Number
1718850-2013-00086
Event Type
Other
Date Received
May 30, 2013
Date of Event
May 3, 2013
Report Date
May 3, 2013
Manufacturer
SORIN GROUP DEUTSCHLAND
Product Code
DWB
PMA / PMN Number
K955038
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SORIN GROUP (B)(4) MANUFACTURES THE S3 DOUBLE HEAD PUMP. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP RECEIVED A REPORT THAT AT THE START OF BYPASS, PRIOR TO CROSS CLAMP, THE CLINICIAN TURNED THE SPEED KNOB OF THE S3 DOUBLE HEAD PUMP AND THE PUMP DID NOT TURN. AFTER SEVERAL SECONDS AN ERROR MESSAGE WAS DISPLAYED. THE CUSTOMER STATED THAT THE ERROR COULD NOT BE CLEARED BY POWER CYCLING THE PUMP UNLESS THE SPEED KNOB WAS RETURNED TO THE ZERO POSITION. THE PUMP WAS CHANGED OUT IN APPROXIMATELY TWO MINUTES. THERE WAS NO REPORT OF PT INJURY. A SORIN SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO EVALUATE THE ISSUE. THE REPRESENTATIVE COMPLETELY DISASSEMBLED THE PUMP, CHECKED ALL CONNECTIONS AND CABLES, REMOVED THE CARD CAGE AND CONTROL BOARDS, AND REASSEMBLED THE PUMP. THE PROBLEM COULD NOT BE REPRODUCED. THE PUMP WAS RUN FOR APPROXIMATELY 50 CYCLES WITH NO FAILURES OR ABNORMALITIES. THE INVESTIGATION IS ON-GOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETED.

Description of Event or Problem · 1

SORIN GROUP RECEIVED A REPORT THAT AT THE START OF BYPASS, PRIOR TO CROSS CLAMP, THE CLINICIAN TURNED THE SPEED KNOB OF THE S3 DOUBLE HEAD PUMP AND THE PUMP DID NOT TURN. AFTER SEVERAL SECONDS AN ERROR MESSAGE WAS DISPLAYED. THE CUSTOMER STATED THAT THE ERROR COULD NOT BE CLEARED BY POWER CYCLING THE PUMP UNLESS THE SPEED KNOB WAS RETURNED TO THE ZERO POSITION. THE PUMP WAS CHANGED OUT. THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
239299 S3 DOUBLE HEAD PUMP CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS DWB SORIN GROUP DEUTSCHLAND 10-65-00 NA

Patients

Seq Age Sex Outcome Treatment
1 67 YR