FDA Adverse Event Injury Summary report: N

SIGN IM NAIL

MDR report key: 3143088 · Received May 29, 2013

Report

Report Number
3034525-2013-00025
Event Type
Injury
Date Received
May 29, 2013
Date of Event
May 24, 2013
Report Date
May 28, 2013
Manufacturer
SIGN FRACTURE CARE INTERNATIONAL
Product Code
HSB
PMA / PMN Number
K022632
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

REVIEWED SURGICAL DATABASE REPORT OF A BROKEN NAIL. REVIEWED SURGEONS REPORT OF A BROKEN IM NAIL DUE TO A NON-UNION. REVIEWED PATIENT X-RAYS WHICH SHOWED A SIGN NAIL BROKEN AT THE NON-UNION AND THE PROXIMAL DISTAL SLOT. THE NON-UNION ALLOWED THE FRACTURE TO SHIFT AND APPLY STRESS TO THE DISTAL END OF THE NAIL AND AT THE SITE OF THE NON-UNION. THE SURGEON PREFORMED AN EXCHANGE NAIL SURGERY TO REPAIR THE FRACTURE AND REPLACE THE IM NAIL. NO INDICATION OF PRODUCT DEFECT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236217 SIGN IM NAIL INTRAMEDULLARY ROD HSB SIGN FRACTURE CARE INTERNATIONAL SIGN IM NAIL UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 24 YR Hospitalization