ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-02745
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- March 27, 2013
- Report Date
- May 20, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: (B)(4) 2013. INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RETURNED WITH THE TISSUE PAD MELTED AND PARTIALLY DETACHED. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND IT DID ACTIVATE DURING FUNCTIONAL TESTING. PROBABLE CAUSES OF TISSUE PAD DAMAGE ARE APPLYING PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. AVOID ACTIVATING THE BLADE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO PREVENT DAMAGE TO THE TISSUE PAD.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE TEFLON PAD FELL OFF INTO THE PATIENT. IT WAS RETRIEVED AND WILL BE RETURNED WITH THE DEVICE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO OTHER DETAILS OF THE EVENT ARE KNOWN. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 243515 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK | J91G4W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR, HANDPIECE |