ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2013-02741
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 22, 2013
- Report Date
- May 21, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). DATE SENT: (B)(4) 2013. INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). DEVICE A WAS RETURNED IN GOOD PHYSICAL CONDITION; THE TISSUE PAD WAS ATTACHED TO THE CLAMP ARM. THE TISSUE PAD OF THE DEVICE WAS IN GOOD PHYSICAL CONDITION. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THE DEVICE WAS TESTED WITH A GENERATOR AND WAS FUNCTIONAL, IN ADDITION, THE DEVICE ACTIVATED WITH BOTH MAX AND MIN BUTTONS. THIS REPORT IS NOT INTENDED TO DENY THAT YOU EXPERIENCED A PROBLEM WITH THE DEVICE. THERE MAY HAVE BEEN OTHER CIRCUMSTANCES OR ISSUES THAT OCCURRED DURING THE USE OF THE DEVICE THAT WE WERE UNABLE TO DUPLICATE DURING OUR LABORATORY ANALYSIS. DEVICE B WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND RETURNED WITH THE DEVICE. THE REMAINING BLADE PORTION WAS SCRATCHED ¿ EVIDENCE OF CONTACT WITH METAL IN OR OUT OF THE OPERATIVE FIELD. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON THE GEN11 GENERATOR, THE ¿INSTRUMENT ERROR¿ ALERT WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE STOP ACTIVATING AND DISPLAY AN INSTRUMENT ERROR SCREEN IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN YELLOW ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿ OR ¿BLADE ERROR DETECTED¿ FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE BLADE WAS BROKEN AND THE PAD WAS BROKEN. NO PATIENT CONSEQUENCES REPORTED. TWO DEVICES WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244656 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |