FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3142782 · Received June 3, 2013

Report

Report Number
2531779-2013-07499
Event Type
Malfunction
Date Received
June 3, 2013
Report Date
May 7, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 07/24/2013 -DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE BOLUS BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED HARD PRESSES TO ELICIT A RESPONSE. THE BOLUS BUTTON HAD A HOLE ON THE COVER. REMOVED BUTTON COVER AND FOUND THE INTERNAL PLUG CONTACT IS MISALIGNED AND CONTAMINATION IS PRESENT.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (AUDIO BOLUS BUTTON TACTILE CHANGE) ISSUE. THE REPORTER STATED THE AUDIO BOLUS BUTTON WAS DIFFICULT TO PRESS, REQUIRED MULTIPLE PRESSES, AND AT TIMES WOULD CAUSE A BOLUS TO NOT BE DELIVERED. THE REPORTER ALLEGED THE BLOOD GLUCOSE (BG) LEVEL WENT TO 18 MMOL/L WITH NO SIGNS OR SYMPTOMS OF HYPERGLYCEMIA. THE REPORTED BG WAS NOT INDICATIVE OF A SERIOUS INJURY AND DOES NOT MEET THE ANIMAS CRITERIA OF AN ADVERSE EVENT. THE REPORTER NOTED AN INDENT IN THE AUDIO BOLUS BUTTON DUE TO HARD BUTTON PRESSES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244538 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 42 YR