FDA Adverse Event Malfunction Summary report: N

HARMONIC SYNERGY* CURVED BLADE

MDR report key: 3142777 · Received June 3, 2013

Report

Report Number
3005075853-2013-02739
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 20, 2013
Report Date
May 21, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
LFL
PMA / PMN Number
K012176
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DATE SENT: (B)(4) 2013. INFORMATION NOT ASKED FOR BUT UNKNOWN OR PROVIDED DURING INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED WITH THE DISTAL TIP OF THE BLADE BROKEN OFF AND RETURNED. THE REMAINING BLADE PORTION WAS SCRATCHED. THE DEVICE WAS FUNCTIONALLY TESTED WITH A GENERATOR. DURING FUNCTIONAL TESTING ON GEN04 AN ERROR CODE 5 WAS DISPLAYED. A PROBABLE CAUSE OF THE DEVICE TO STOP ACTIVATING AND DISPLAY AN ERROR CODE 5 IS BLADE DAMAGE. PROBABLE CAUSES OF BLADE DAMAGE, INCLUDING BREAKAGE, ARE EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE, BLADE CONTACT WITH OTHER DEVICES, STAPLES OR CLIPS DURING THE PROCEDURE OR USING ANY MEANS OTHER THAN THE BLADE WRENCH TO ATTACH OR DETACH THE BLADE. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN AN ERROR CODE 5 OR BLADE LOCKOUT LATER IN THE PROCEDURE, AND CONTINUED USAGE CAN RESULT IN A BROKEN BLADE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CONIZATION, AN UNEXPECTED NOISE WAS HEARD. WHEN THE DEVICE WAS CHECKED OUTSIDE THE PATIENT, THE BLADE WAS BROKEN OFF OUTSIDE THE PATIENT. THE DEVICE WAS USED FOR THE UTERUS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. ONE DEVICE WILL BE RETURNING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244717 HARMONIC SYNERGY* CURVED BLADE INSTRUMENT, ULTRASONIC SURGICAL LFL ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR AND HANDPIECE