FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3142746 · Received May 24, 2013

Report

Report Number
1720753-2013-06409
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
May 3, 2013
Report Date
May 24, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM HAD A FILAMENT ERRORS, COMMUNICATION PROBLEMS AND DATA ERRORS AND WOULD NOT BOOT UP. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
232302 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1