FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3142746
·
Received May 24, 2013
Report
- Report Number
- 1720753-2013-06409
- Event Type
- Malfunction
- Date Received
- May 24, 2013
- Date of Event
- May 3, 2013
- Report Date
- May 24, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR INTERFACE BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM HAD A FILAMENT ERRORS, COMMUNICATION PROBLEMS AND DATA ERRORS AND WOULD NOT BOOT UP. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232302 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |