FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3142733 · Received June 3, 2013

Report

Report Number
2210968-2013-06405
Event Type
Malfunction
Date Received
June 3, 2013
Date of Event
May 15, 2013
Report Date
May 15, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. THIS IS ONE OF THREE MEDWATCHES BEING SUBMITTED AS THREE DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-06404 AND MEDWATCH 2210968-2013-06406. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE MAIN HOUSING WAS RECEIVED IN A DISASSEMBLED CONDITION. FUNCTIONAL TESTS INVOLVING THE MAIN HOUSING COULD NOT BE PERFORMED. UPON EVALUATION, THE TRIGGER OPERATED AS INTENDED.IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A MYOMECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE SURGEON MORCELLATED THE 90 G MYOMA AND IT WORKED PROPERLY. AFTER THAT, THE 860G MYOMA WAS MORCELLATED WITH THE SAME DEVICE. AFTER TEN TO FIFTEEN MINUTES, THE DEVICE WOULD NOT ACTIVATE. ANOTHER LIKE DEVICE WAS USED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
245058 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT217975

Patients

Seq Age Sex Outcome Treatment
1