FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3142641 · Received May 24, 2013

Report

Report Number
2518422-2013-00951
Event Type
Malfunction
Date Received
May 24, 2013
Date of Event
January 4, 2011
Report Date
January 4, 2011
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED TO POWER ON. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE'S DISPLAY WAS BLANK. THE DEVICE'S INVERTER BOARD WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MANUFACTURER'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
231410 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1