FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3142622
·
Received June 3, 2013
Report
- Report Number
- 1416980-2013-14080
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 13, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. THIS IS AN ANCILLARY SERVICE EVENT. THE ROOT CAUSE OF ALARM F-63 WAS DETERMINED TO BE DAMAGE TO THE CENTRAL PROCESSING UNIT BOARD (CPU). TO CORRECT THE CONDITION, THE CPU BOARD WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
DURING THE PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE AN F-63 ALARM.¿ NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244846 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |